How stiff must a fixture be to resemble a fixed gap?
Give a 1000 N/mm stack 1 mm free growth. Compare fixture stiffness 100, 1000 and 10000 N/mm with the rigid limit; use no preload or gap.
The fixture opens under force, consuming some of the free growth that would otherwise compress the stack. The force ratio K/(K+kc) quantifies that loss. Calling the displacement setting fixed does not remove compliance from the actual hardware.
Connect the physical quantities
The relationship used in this example
A swelling test does not measure an intrinsic force independently of its restraint. A freely expanding specimen, a force-controlled specimen and a nominally fixed-gap specimen follow different mechanical paths. Describe both the specimen and its surroundings before comparing curves from different fixtures.
Work through the example
At K = 1000 N/mm, total compliance is 0.001+0.001 = 0.002 mm/N.
Force is 1/0.002 = 500 N. Fixture opening and relative stack compression are each 0.5 mm.
At K = 10000 N/mm, force becomes 1/0.0011 = 909.091 N; check 0.090909+0.909091 = 1 mm.
Full example conditions and reference values
Use a synthetic stack with kc = 1000 N/mm and free growth Δf = 1 mm, with no gap or initial preload. For K = 100, 1000 and 10000 N/mm, forces are 90.909, 500 and 909.091 N. Fixture openings are 0.909091, 0.5 and 0.090909 mm. The rigid-fixture force is 1000 N, so even a fixture ten times stiffer than the stack gives a force 9.09% below the rigid limit.
Compare the numbers before judging the result
| K/kc | Force (N) | Force / rigid (%) | Fixture opening (mm) |
|---|---|---|---|
| 0.1 | 90.90909 | 9.090909 | 0.9090909 |
| 1 | 500 | 50 | 0.5 |
| 10 | 909.0909 | 90.90909 | 0.09090909 |
| 100 | 990.099 | 99.0099 | 0.00990099 |
A fixture one hundred times stiffer still has finite compliance; its effect is quantified rather than named away.
What the result means for a model
Measure the opening between the actual compression faces where possible. Choose the required stiffness relative to the acceptable measurement bias and stack compliance; the ratios here are illustrative, not a universal fixture specification.
Another application: How rigid is a fixed-gap fixture?
In the linear no-pad model, the force-to-rigid-limit ratio is K/(K+kc). For K = 10kc it is 90.91%; K = 100kc gives 99.01%. “Fixed gap” in a test description is therefore not enough to establish negligible fixture deformation.
Further modeling details
Start with the controlled quantity
A free-growth test allows displacement with negligible incremental force. A force-controlled test adjusts position to retain a target force, while a gap-controlled test attempts to retain a measured spacing. State which sensor closes the control loop: actuator position and local platen separation are not equivalent when the load train deforms.
Build a displacement budget
Free growth is accommodated by fixture opening and compression of the stack relative to its free state. Their sum must equal Δf. Force equilibrium gives the same force in these series components. The formula follows by adding their compliances; it is not an empirical battery law.
Define what rigid means for the purpose
For this model, rigid-limit force error is kc/(K+kc). Requiring an error smaller than a chosen 5% means K must exceed 19kc. This is an illustrative decision derived for one scalar output, not a universal test-standard requirement. Nonlinear specimen stiffness changes the relevant ratio as loading progresses.
Separate preload from subsequent growth
The displayed example starts without preload. If a linear system is already assembled at F0 and maintains contact, write the same compatibility for increments and add ΔF to F0. If contact opens, a pad densifies or a controller holds force constant, that incremental spring model must be changed.
Design a useful comparison
Record force, local thickness or gap, temperature and the electrical state on a synchronized time base. Repeat selected states with a second known restraint stiffness. A model intended for a module should be checked under a relevant constraint, rather than matching one rigid fixture and assuming every enclosure behaves identically.
Formula and symbol reference
Open the complete formula reference
| Symbol | Meaning | Unit / convention |
|---|---|---|
| Δf, Δh, Δc | Free growth, fixture opening, compression relative to free state | mm |
| kc, K | Stack and fixture axial stiffness | N/mm |
| F | Compressive force magnitude; incremental if linearized about preload | N |
When this example stops applying
One-dimensional, linear, uniform-temperature teaching model. No gas-pressure distribution, plate bending, friction, feedback-controller dynamics or viscoelasticity is resolved. Example stiffnesses are synthetic.
Check your interpretation
Can a higher force in a second fixture alone prove the cell became stiffer?
Show the reasoning
No. A stiffer restraint alone increases force for the same stack stiffness and free growth. Compare the complete boundary and measurement conditions before attributing a difference to material behavior.
Sources and calculation scope
- MIT OCW · Mechanical behavior and elasticity foundations
- Instron · Universal Testing Machine Compliance (measurement background)
The questions and comparison tables are independently authored teaching examples. Numbers illustrate the stated models; they are not measured cell/pack data or commercial-solver validation.