Theory / Verification & Correlation

Machine Compliance & Local Displacement

Remove only the measured load-train contribution and understand how uncertainty affects inferred stiffness.

Theory · Educational notes

What did the displacement sensor actually measure?

At 1000 N, the crosshead moves 0.15 mm. An independent fixture estimate is 0.00005 mm/N. Remove fixture motion before calculating specimen stiffness.

Series deformations add, so the measured motion is not automatically specimen motion. Correcting it is a subtraction of uncertain quantities. When specimen motion is small, even a modest fixture uncertainty can dominate the corrected result.

Connect the physical quantities

The relationship used in this example

us=um−CfFu_s=u_m-C_fF

A crosshead travels to deform both the specimen and the load train. A force–crosshead curve therefore identifies system compliance unless those contributions are separated. A local displacement sensor can reduce the ambiguity, but its reference points and calibration still matter.

Work through the example

Step 1

Fixture displacement: 1000×0.00005 = 0.05 mm.

Step 2

Specimen displacement: 0.15−0.05 = 0.10 mm; stiffness: 1000/0.10 = 10000 N/mm.

Step 3

Uncertainty contributions are 0.002, 0.005 and 0.0005 mm; root-sum-square is 0.005408 mm.

Full example conditions and reference values

At F = 1000 N, let measured crosshead displacement be 0.15 mm and separately estimated fixture compliance be 0.00005 mm/N. The fixture contributes 0.05 mm, leaving 0.10 mm specimen displacement. Apparent system stiffness is 6666.67 N/mm while corrected specimen stiffness is 10000 N/mm. With independent standard uncertainties of 0.002 mm in crosshead displacement, 0.000005 mm/N in fixture compliance and 10 N in force, propagated displacement uncertainty is 0.005408 mm.

Machine Compliance & Local Displacement — synthetic teaching example
Original Jaylab analytical illustration with synthetic inputs. No measured battery or fixture data are shown.

Compare the numbers before judging the result

Compare the calculated values
Corrected motion (mm)Absolute uncertainty (mm)Relative (%)
0.10.0054083275.408327
0.010.00540832754.08327

The second row holds the same illustrative absolute uncertainty, showing the subtraction problem for a much stiffer specimen.

What the result means for a model

Use local sensing or improve fixture characterization rather than reporting extra decimal places. The example propagation assumes independent uncertainties; shared calibration errors require covariance terms.

Another application: Subtraction can magnify uncertainty

The corrected specimen displacement is 0.10 mm with standard uncertainty about 0.005408 mm, or 5.41%. If a stiffer specimen produced only 0.01 mm corrected motion with similar absolute uncertainty, the relative uncertainty would be about 54%. The subtraction is then poorly resolved.

Further modeling details

Draw the measurement loop

Mark the two points between which each displacement is measured. A platen-to-platen reading includes material between the platens; an actuator encoder includes more of the machine. Neither should be renamed specimen strain without dividing the appropriate gauge displacement by its reference length.

Characterize the same load train

A compliance correction should represent the actual grips, couplings and fixture configuration over the load range. A dummy is not infinitely rigid: account for its own deformation where relevant. Seating and friction may make the correction nonlinear or history-dependent, so a single slope can be insufficient.

Subtract displacement before inferring modulus

At a given force, subtract fixture deformation from the total. For a uniform specimen, convert the resulting displacement with E = FL/(Aus). Do not subtract fixture stiffness directly from measured stiffness: compliances add in series, not stiffnesses.

Track uncertainty and subtraction sensitivity

The uncertainty equation assumes independent errors and first-order propagation. Correlated calibration errors require covariance terms. When the corrected specimen displacement is a small difference between two large values, relative uncertainty can become large even if the raw signals look smooth.

Use a diagnostic stop rule

If Cm−Cf is zero or negative within uncertainty, do not report an infinite or negative specimen modulus. Recheck calibration, units, reference locations and loading history. Compare against a local displacement measurement or a more compliant specimen before refining the constitutive model.

Formula and symbol reference

Open the complete formula reference
um=us+uf,uf=CfFu_m=u_s+u_f,\qquad u_f=C_fF
us=um−CfFu_s=u_m-C_fF
Cm=Cs+Cf,ks=1Cm−CfC_m=C_s+C_f,\qquad k_s=\frac{1}{C_m-C_f}
sus2≈sum2+(FsCf)2+(CfsF)2s_{u_s}^2\approx s_{u_m}^2+(Fs_{C_f})^2+(C_fs_F)^2
SymbolMeaningUnit / convention
um, us, ufMeasured crosshead, specimen and fixture displacementmm
Cf, Cm, CsFixture, measured-system and specimen compliancemm/N
sIndependent small standard uncertainties in the displayed approximationmatching variable units

When this example stops applying

Linear series compliance and small independent uncertainty approximation. No standard-compliant machine calibration or battery material identification is claimed. Nonlinear seating, hysteresis and thermal drift need separate treatment.

Check your interpretation

Can fixture stiffness simply be subtracted from measured stiffness?

Show the reasoning

No. For the stated series system, subtract compliances or subtract displacement at the same force. Invert the remaining positive specimen compliance only afterward.

Sources and calculation scope

The questions and comparison tables are independently authored teaching examples. Numbers illustrate the stated models; they are not measured cell/pack data or commercial-solver validation.